Prologix GPIB-USB Controller User Manual
9/15/2009 12
++mode 0 Switch to DEVICE mode
++mode Query current mode
8.11. read
This command can be used to read data from an instrument until:
EOI is detected or timeout expires, or
A specified character is read or timeout expires, or
Timeout expires
Timeout is set using the read_tmo_ms command and applies to inter-character delay, i.e.,
the delay since the last character was read. Timeout is not be confused with the total
time for which data is read.
SYNTAX: ++read [eoi|<char>] where <char> is a decimal value less than 256
MODES AVAILABLE: CONTROLLER
EXAMPLES:
++read Read until timeout
++read eoi Read until EOI detected or timeout
++read 10 Read until LF (ASCII 10) is received or timeout
8.12. read_tmo_ms
This command specifies the timeout value, in milliseconds, to be used in the read
command and spoll command. Timeout may be set to any value between 1 and 3000
milliseconds.
SYNTAX: ++read_tmo_ms <time> where <time> is decimal value between 1 and
3000
MODES AVAILABLE: CONTROLLER
8.13. rst
This command performs a power-on reset of the controller. The process takes about 5
seconds. All input received over USB during this time are ignored.
SYNTAX: ++rst
MODES AVAILABLE: CONTROLLER, DEVICE